D.O.R.C. ZX-1 Mini Zoom Interferometer with electronic zoom measures the surface topography of connectors or fiber endface, bare fiber holder chuck with x.y.z and tilt micrometers.
Pictured without chuck, but chuck is included. 1.5mm, 2.5mm and FC chucks available. Please request chuck type at time of purchase.
Manufacturer DescriptionKey Benefits
Product Information:
Direct Optical Research Company specializes in the design and manufacturing of non-contact interferometers for the optical fiber and connector industry. DORC's range of ZX-1 Zoom Interferometers and Phase Measuring System (PMS) software has been specifically designed to meet the ever-increasing need for accurate, automated, high volume testing of ferrules, ferrule blanks, connector assemblies, pigtails, patchcords, array connectors and optical fiber cleaves.
DORC's implementation of the Michelson Interferometer incorporates an electronically controlled zoom objective. Fringes created by introducing an internal reference path can be analyzed to measure the surface topography of a connector or fiber endface. The fully enclosed vibration and contamination resistant Michelson configuration eliminates the requirement for the delicate surface of the sample under test to be in physical contact with the measuring equipment. Being non-contact not only avoids contamination of both the sample and interferometer, but also removes the possibility of contamination induced measurement errors.
| Radius of Curavture |
Reproducibility - One Sigma: 0.25% Repeatibility - One Sigma: 0.10% |
| Apex Offset |
Reproducibility - One Sigma: 2.0 µm Repeatibility - One Sigma: 0.5 µm |
| Fiber Height |
Reproducibility - One Sigma: 1.5 nm Repeatibility - One Sigma: 1.0 nm |
| Radius of Curvature | 3mm-Flat |
| Apex Offset | 0-500 µm (value calculated if outside of captured frame) |
| Fiber Height | ± 6 µm (with extended range option) |
| Principle of Operation | Michelson Interferometer (non-contact system) |
| Measurement Speed | Less than 1 second (standard range) |
| Magnification |
X85 - X550 for ZX-1 mini using 12" monitor X60 - X400 for a ZX-1 mini PMS using a 15" XGA monitor and the digital zoom turned off |
| Operating Wavelengths | 665 nm (and 545 nm with extended range) |
| Power Requirements for ZX-1 |
100-120V AC 50/60Hz (USA Version) 100-120V and 210-250V AC 50/60Hz auto sensing (International Version) |